Bil Micro Meter Auto-3




The AUTO-3 is an advanced model for quickly measuring the total bilirubin in serum without being affected by an influence of the light.

The Auto-3 features ease of operation, high stability, compact size and low power consumption.

The conventional principle of measuring maximum absorption of bilirubin at the wavelength of 455 nm and subtracting the turbidity caused by hemolysis and contamination at the wavelength of 575 nm is followed in this model, but a longer life and higher performance are secured by the adoption of new technology such as a white LED for light source and a 16 bit microprocessor.

Features

Easier operation
Thanks to use of advanced electronic technology such as a microprocessor, there is no need for frequent adjustment.
Auto-zero is automatically done by push of Auto-zero button.
An alarm signals abnormalities such as high concentrations or high hemolytic hemoglobin etc.

High stability and reliability
Dual wavelength analysis corrects the influences of hemolysis and turbidity.
A long-life white LED as a light source provides the stable measurement.

Compact size and light-weight
Weight is about 2.0kg and it is easy to carry.

Low power consumption
Only 15VA of power consumption saves running cost.

Specifications

Measuring range 0 to 30 mg/dl (total bilirubin)
Response time 3 seconds
Correcting hemolysis 0 to 250 mg/dl
Concentration alarm Indicates 999 for 31 mg/dl or more
Turbidity alarm Turbidity of 250 mg/dl or more
Filters 455nm & 575nm
0-Adjustment Automatically adjusted by pushing a button
Sample blood volume 50 to 60µl (whole blood)
Cell Designated capillary tube
Display 3-digit 7-segment red LED
Light source White LED
Dimensions 230(w)x200(D)x100(H) mm
Weight Approx. 2.0 kg
Power supply AC 90-240V 50/60Hz 15VA